DEMERS AMBULANCES MXP153 DODGE RAM 2020: EQUIPMENT:ELECTRICAL

General Information

NHTSA Campaign ID Number: 21V740000

Make Model Year
DEMERS AMBULANCES MXP153 DODGE RAM 2020
Date Owners Notified Date Received by ODI Date Added to Database
2021-10-06 2021-09-23 2021-09-24

Manufacturers Involved: Braun Ambulances

Component: EQUIPMENT:ELECTRICAL

Potential Number Of Units Affected: 9

Report Initiator: MFR MFR Report

Summary

Braun Ambulances is recalling certain 2020 Demers Ambulances MXP153 Ford F-Series Gas, MXP153 Dodge Ram, and 2020-2021 MXP153 Ford F-Series Diesel emergency vehicles. The medical oxygen tube and electrical wires may become damaged due to insufficient clearance from the ceiling spacers.

Consequence

Damage to the wires and medical oxygen tube increases the risk of a fire.

Remedy

Dealers will remove the ceiling panels and spacers, replace wires and oxygen network tubing, and create a routing channel inside the insulation board for all wires, free of charge. Owner letters were mailed on October 6, 2021. Owners may contact Braun Ambulances' customer service at 1-877-344-9990.

Notes

Owners may also contact the National Highway Traffic Safety Administration Vehicle Safety Hotline at 1-888-327-4236 'TTY 1-800-424-9153', or go to www.nhtsa.gov.